Teledyne 3000 Ultra Trace - PPB oxygen analyzer User Manual

Page 55

background image

Ultra Trace Oxygen Analyzer

Operation 4

4-21

Teledyne Analytical Instruments

Use the

∆∇

arrow keys to enter the upper value of the range (all ranges

begin at 0). Repeat for each range you want to set. Press

Enter

to accept the

values and return to

Analyze

mode. (See note below.)

Note:

The ranges must be increasing from low to high, for example, if range
1 is set as 0–500 ppb and range 2 is set as 0–10 ppm, range 3 cannot
be set as 0– 5 ppm since it is lower than range 2.

Ranges, and alarms, are set in ppb or ppm units depending on concen-

tration. All concentration-data outputs change from ppb units to ppm when
the concentration is above 1.0 ppm. Range Low (L) is always a ppb range
and cannot be set higher than 1000 ppb nor lower than 10 ppb. Ranges
medium (M) and High (H) can only be set in ppm. The Medium (M) range
can be set between 1 and 10 ppm, while the high (H) range can be set
between 10 and 1000 ppm.

4.7.2 Fixed Range Analysis

The autoranging mode of the instrument can be overridden, forcing the

analyzer DC outputs to stay in a single predetermined range.

To switch from autoranging to fixed range analysis, enter the range

function by pressing the

Range

button on the front panel.

Use the < > arrow keys to move the blinking over

AUTO.

Use the

∆∇

arrow keys to switch from

AUTO to FX/L, FX/M, or FX/H to

set the instrument on the desired fixed range (low, medium, or high).

L—250ppb M—1 ppm

H—10 ppm Mode—FX/L

or

L—250 ppb M—1 ppm

H—10 ppm Mode—FX/M

or

L—250 ppb M—1 ppm

H—10 ppm Mode—FX/H

Press

Escape

to re-enter the

Analyze

mode using the fixed range.

NOTE: When performing analysis on a fixed range, if the oxygen concentra-

tion rises above the upper limit (or default value) as established by the
operator for that particular range, the output saturates at 1 V dc (or 20
mA). However, the digital readout and the RS-232 output continue to
read the true value of the oxygen concentration regardless of the
analog output range.