beautypg.com

Teledyne 3000 Ultra Trace - PPB oxygen analyzer User Manual

Page 49

background image

Ultra Trace Oxygen Analyzer

Operation 4

4-15

Teledyne Analytical Instruments

This menu allows the operator to set the time the analyzer should be

held in the AUTO span mode, after the readings of the analyzer settle. Five
minutes is the default, but it could be adjusted anywhere from 1 to 60 min-
utes by using the UP or DOWN keys.

Press

Enter

to move to the next screen.

Span Val: 008.00 ppm

Span Mod #

Use the

∆∆∆∆∆∇

arrow keys to enter the oxygen-concentration mode

(999.99 ppm is maximum value of span gas allowed). Use the < > arrow
keys to blink the digit you are going to modify. Use the

∆∆∆∆∆∇

arrow keys

again to change the value of the selected digit. When you have finished
typing in the concentration of the span gas you are using, press

Enter

to

begin the Span calibration.

#### ppm Span

Slope=#### ppm/s

The beginning span value is shown in the upper left corner of the

display. As the span reading settles, the screen displays and updates informa-
tion on

Slope. Spanning automatically ends when the span output corre-

sponds, within tolerance, to the value of the span gas concentration. Then the
instrument automatically returns to the analyze mode.

4.4.2.2

Manual Mode Spanning

Press

Span

to start the

Span

function. The screen that appears allows

you to select whether the span calibration is to be performed automatically or
manually.

Span: Settling:MAN

For Next

Use the

∆∇

keys to toggle between

AUTO and MAN span settling. Stop

when

MAN appears, blinking, on the display. Press

Enter

to move to the

next screen.

Press

Enter

to move to the next screen.

Calib. Holding time

Cal hold: 5 min

This menu allows the operator to set the time the analyzer should be

held in the AUTO span mode. It does not have any effect in the MANual
mode. Just press Enter key to continue.