Teledyne 3000 Ultra Trace - PPB oxygen analyzer User Manual

Page 63

background image

Ultra Trace Oxygen Analyzer

Maintenance 5

5-5

Teledyne Analytical Instruments

5.3

Fuse Replacement

1. Place small screwdriver in notch, and pry cover off, as shown in

Figure 5-2.

Figure 5-2: Removing Fuse Block from Housing

2. To change between American and European fuses, remove the

single retaining screw, flip Fuse Block over 180 degrees, and
replace screw.

3. Replace fuse as shown in Figure 5-3.

4. Reassemble Housing as shown in Figure 5-2.

American Fuses

European Fuses

Figure 5-3: Installing Fuses