Teledyne 3000 Ultra Trace - PPB oxygen analyzer User Manual

Page 12

background image

1-6

1 Introduction

Model Ultra Trace 3000

Teledyne Analytical Instruments

RS-232 Port

Serial digital concentration signal output
and control input.

Remote Probe

Used in the Ultra Trace3000 for
controlling external solenoid valves
only.

Remote Span/Zero

Digital inputs allow external control of
analyzer calibration.

Calibration Contact

To notify external equipment that
instrument is being calibrated and
readings are not monitoring sample.

Range ID Contacts

Four separate, dedicated, range relay
contacts. Low, Medium, High, Cal.

Network I/O

Serial digital communications for local
network access. For future expansion.
Not implemented at this printing.

Note: If you require highly accurate Auto-Cal timing, use external

Auto-Cal control where possible. The internal clock in the
Model Ultra Trace 3000 is accurate to 2-3 %. Accordingly,
internally scheduled calibrations can vary 2-3 % per day.