beautypg.com

Teledyne 3000 Ultra Trace - PPB oxygen analyzer User Manual

Page 27

background image

Ultra Trace Oxygen Analyzer

Installation 3

3-5

Teledyne Analytical Instruments

3.3.2.2

50-Pin Equipment Interface Connector

Figure 3-4 shows the pin layout of the Equipment Interface connector.

The arrangement is shown as seen when the viewer faces the rear panel of
the analyzer. The pin numbers for each input/output function are given
where each function is described in the paragraphs below.

Figure 3-4: Equipment Interface Connector Pin Arrangement

Analog Outputs: There are four DC output signal pins—two pins per

output. For polarity, see Table 3-1. The outputs are:

0–1 V dc % of Range: Voltage rises linearly with increasing oxygen, from

0 V at 0 ppm to 1 V at full scale ppm. (Full scale =
100% of programmable range.)

0–1 V dc Range ID:

0.25 V = Low Range, 0.5 V = Medium Range,
0.75 V = High Range, 1 V = Air Cal Range.

4–20 mA dc % Range: Current increases linearly with increasing oxygen,

from 4 mA at 0 ppm to 20 mA at full scale ppm.
(Full scale = 100% of programmable range.)

4–20 mA dc Range ID: 8 mA = Low Range, 12 mA = Medium Range, 16

mA = High Range, 20 mA = Air Cal Range.

Table 3-1: Analog Output Connections

Pin

Function

3

+ Range ID, 4-20 mA, floating

4

– Range ID, 4-20 mA, floating

5

+ % Range, 4-20 mA, floating

6

– % Range, 4-20 mA, floating

8

+ Range ID, 0-1 V dc

23

– Range ID, 0-1 V dc, negative ground

24

+ % Range, 0-1 V dc

7

– % Range, 0-1 V dc, negative ground

Alarm Relays: The nine alarm-circuit connector pins connect to the

internal alarm relay contacts. Each set of three pins provides one set of Form
C relay contacts. Each relay has both normally open and normally closed