4 Operation

Model Ultra Trace 3000

4-22

Teledyne Analytical Instruments

4.8

The

Analyze

Function

Normally, all of the functions automatically switch back to the

Analyze

function when they have completed their assigned operations. Pressing the

Escape

button in many cases also switches the analyzer back to the

Ana-

lyze

function. Alternatively, you can press the

Analyze

button at any time

to return to analyzing your sample.

4.9

Signal Output

The standard Model Ultra Trace 3000 Oxygen Analyzer is equipped

with two 0–1 V dc analog output terminals accessible on the back panel

(one concentration and one range ID), and two isolated 4–20 mA dc

current outputs (one concentration and one range ID).

See Rear Panel in Chapter 3, Installation, for illustration.

The signal output for concentration is linear over the currently selected

analysis range. For example, if the analyzer is set on range that was defined
as 0–100 ppm O

2

, then the output would be:

Voltage Signal

Current Signal

ppm O

2

Output (V dc)

Output (mA dc)

0

0.0

4.0

10

0.1

5.6

20

0.2

7.2

30

0.3

8.8

40

0.4

10.4

50

0.5

12.0

60

0.6

13.6

70

0.7

15.2

80

0.8

16.8

90

0.9

18.4

100

1.0

20.0

The analog output signal has a voltage which depends on the oxygen

concentration AND the currently activated analysis range. To relate the
signal output to the actual concentration, it is necessary to know what range
the instrument is currently on, especially when the analyzer is in the
autoranging mode.