Ultra Trace Oxygen Analyzer

Operation 4

4-3

Teledyne Analytical Instruments

4.3

The

System

Function

The subfuctions of the

System

function are described below. Specific

procedures for their use follow the descriptions:

Auto-Cal: Used to define an automatic calibration sequence
and/or start an Auto-Cal.

PSWD: Security can be established by choosing a 5 digit
password (

PSWD) from the standard ASCII character set. (See

Installing or Changing the Password, below, for a table of
ASCII characters available.) Once a unique password is assigned
and activated, the operator MUST enter the UNIQUE password
to gain access to set-up functions which alter the instrument's
operation, such as setting the instrument span or zero setting,
adjusting the alarm setpoints, or defining analysis ranges.

Figure 4-1: Hierarchy of Functions and Subfunctions

Set Instrument

Span

Set Instrument

Zero

Set Alarm

Setpoints

Confrigure Mode

of Alarm

Operation

Define Analysis

Ranges

Perform Oxygen

Analysis of

the Sample

ANALYZE

SYSTEM

SPAN

ZERO

ALARMS

RANG

-

TRAK/HLD

Perform

Self-Diagnostic

Test

Initiate

Automatic

Calibration

Set Password

Logout

Neg