beautypg.com

Teledyne 3000 Ultra Trace - PPB oxygen analyzer User Manual

Page 25

background image

Ultra Trace Oxygen Analyzer

Installation 3

3-3

Teledyne Analytical Instruments

3.3

Rear Panel Connections

Figure 3-3 shows the Model Ultra Trace 3000 rear panel. There are

ports for gas inlet and outlet, power, communication, and both digital and
analog concentration output.

Figure 3-3: Rear Panel of the Model Ultra Trace 3000

3.3.1 Gas Connections

The unit is manufactured with

1

/

4

inch VCR fittings. For a safe connec-

tion:

SAMPLE IN: In the standard model, gas connections are made at the

SAMPLE IN

and

EXHAUST OUT

connections. Calibration gases must be

Tee'd into the Sample inlet with appropriate valves. A VCR fitting is pro-
vided for the inlet connection.

The inlet gas pressure should be reasonably regulated. Pressures be-

tween 1 and 50 psig are acceptable as long as the pressure, once established,
will keep the front panel flowmeter reading in an acceptable range (0.5 to 2.0
SLPM). For non-pressurized sample or very low pressure, (less than 1 psig)
vacuum service plumbing is recommended. (See next section: Vacuum
Service).

Te le d y n e A n aly t ic a l In s tru m e n ts