Ultra Trace Oxygen Analyzer

Maintenance 5

5-3

Teledyne Analytical Instruments

5.2.3 Removing the Micro-Fuel Cell

The Micro-Fuel cell is located inside the stainless steel cell block behind

the front panel (see Figure 5-1). To remove an existing cell:

1. Remove power to the instrument by unplugging the power cord

at the power source.

2. Open the front panel door by pressing the release button on the

top right corner of the door all the way in with a narrow gauge
tool.

3. With one hand hold the top of the cell block while unscrewing

the plastic ring holder. Once the plastic ring is loose, remove the
top of the cell block.

Figure 5-1: Removing the Micro-Fuel

COLLAR

CELL BLOCK

SENSOR

O-RING

CELL HOLDER