Ultra Trace Oxygen Analyzer

Operation 4

4-1

Teledyne Analytical Instruments

Operation

4.1

Introduction

Once the analyzer has been installed, it can be configured for your

application. To do this you will:

Set system parameters:

Establish a security password, if desired, requiring Operator
to log in.

Establish and start an automatic calibration cycle, if desired.

Calibrate the instrument.

Define the three user selectable analysis ranges, then choose
autoranging or select a fixed range of analysis, as required.

Set alarm setpoints, and modes of alarm operation (latching,
failsafe, etc).

Before you configure your Ultra Trace 3000, these default values are in

effect:

Ranges: LO = 250ppb ppm, MED = 1 ppm, HI = 10 ppm.
Auto Ranging: ON
Alarm Relays: Defeated, Alarm 1 at10.000 ppm, Alarm 2 at 1.000

ppm HI, Not failsafe, Not latching.

Zero: Auto, every 0 days at 0 hours.
Span: Auto, at 008.00 ppm, every 0 days at 0 hours.

If you choose not to use password protection, the default password is

automatically displayed on the password screen when you start up, and you
simply press

Enter

for access to all functions of the analyzer.