RATAmation v2.0 User Manual
Perma Pure LLC
Phone (732) 244-0010
www.permapure.com
Sept. 2012
28
Selecting a line of the grid will copy the data to fields where the info can be edited.
There are two buttons, “OK” and “Undo”, near the editing fields.
“OK” Button - will replace the current grid information with the data currently in the edit
fields. The new data will be placed into the grid but not stored to the configuration file.
“Undo” Button - will discard changes and replace the values with the information in the
grid for the current line being edited.
The following information can be set for each of the calibration gases of the flow drawer.
Cylinder ID – descriptive text for the cylinder of blended gas
Used for reference only
A blended gas can consist of up to 4 components. The type and quantity for each component
must be entered. Below each of the labels Component Gas 1 though 4 is an entry box to enter the
quantity and a selection list of gas types
Quantity -
Quantity of the specific type of gas selected below must be in same units as
analyzer.
Gas Type-
Not Used / SO2 / NOx / CO2 / O2 / CO / THC / TRS / N2
The default value is Not Used.
There are three buttons below the grid.
Cancel –
Restores the information shown in the grid to the last values stored as the
configuration information on a disk drive.
Accept –
Saves the current information contained in the grid to the configuration file stored
on the disk drive
Exit –Closes the Analyzer Input Configuration screen.
The “X” on the screen will also close the screen