Ratamation v2.0 user manual perma pure llc – Perma Pure RATAMATION V2.0 User Manual
Page 26

RATAmation v2.0 User Manual
Perma Pure LLC
Phone (732) 244-0010
www.permapure.com
Sept. 2012
24
The configurations for all channels are displayed in a spreadsheet type grid.
The screen is initially displayed with the stored setup information used by the program in the
main grid and with Chan 1 selected.
Selecting a line of the grid will copy the data to fields where the info can be edited.
There are two buttons, OK and Undo, near the editing fields.
“OK” Button - will replace the current grid information with the data currently in the edit
fields. The new data will be placed into the grid but not stored to the configuration file.
“Undo” Button - will discard changes and replace the values with the information in the
grid for the current line being edited.
The following information can be set for each of the analog inputs of the flow drawer.
Input Type – Unused / Analyzer / Other Data
Chan 1-8
Unused / Other Data
Chan 9-12
The value defaults to Unused.
Unused is not displayed on any screen
Analyzer is displayed on all cal and live data screens
Other Data is displayed on all live data screens
Channel Name – descriptive text to be used for data display
Gas Type-
SO2 6c /NOx 7e/ NOx 20/ CO2 3a/ CO2 20/
O2 3a/ O2 20/ CO 10/ THC 25a
The default value is SO2 6c.
Used only when Input Type is set to Analyzer.
The value is ignored otherwise.
Analyzer SN- descriptive text of the serial number for the analyzer.
Used for reference only
Offset Value – the scaling factor to adjust the zero of the input
This value is subtracted from the scaled input to obtain the final displayed and
recorded value.
Full Scale – Full scale input of the Analyzer or Other Data.
The value is used for display only.