The sub-menu is entered by pressing [Enter] –
• Use the [Up] and [Down] keys to increase or decrease the value for
setting. Press [Enter] to accept the setting and advance to the next
item in the menu
• Press [Mode] to advance to setting of the next parameter or [Esc] to
return to normal weighing
13.5 CALIBRATION SETUP
This menu allows the user to set the calibration parameters.
• Press [Enter] when “CAL SETUP” is displayed to select the calibration
parameters
• The options for each parameter can be scrolled through by using the
[Up] or [Down] key
ENABLE
NO =operator calibration is disabled
YES=operator calibration is enabled
CAL REPORT
On = Enabled. Prints out Calibration
report after successful calibration
OFF = Disabled
TIME CAL
On = Enabled. Select time from 1 to
24 hours. Default setting is 4 hours
OFF = Disabled
TEMP CAL
On = Enabled. If set to On, set the
temperature variation from 0.2 to 4°C
OFF = Disabled
INT CAL
(Only on balances with
internal calibration masses)
YES = Use Internal calibration mass
NO = Use external calibration mass
Press [Mode] to advance to setting of the next menu which is
“PASSCODES” or [Esc] to return to normal weighing.