beautypg.com

5 calibration setup – Nova-Tech PGW Series User Manual

Page 34

background image

Page 34 of 44

The sub-menu is entered by pressing [Enter]

• Use the [Up] and [Down] keys to increase or decrease the value for

setting. Press [Enter] to accept the setting and advance to the next
item in the menu

• Press [Mode] to advance to setting of the next parameter or [Esc] to

return to normal weighing

13.5 CALIBRATION SETUP

This menu allows the user to set the calibration parameters.

• Press [Enter] when “CAL SETUP” is displayed to select the calibration

parameters

• The options for each parameter can be scrolled through by using the

[Up] or [Down] key

ENABLE

NO =operator calibration is disabled
YES=operator calibration is enabled

CAL REPORT

On = Enabled. Prints out Calibration
report after successful calibration
OFF = Disabled

TIME CAL

On = Enabled. Select time from 1 to
24 hours. Default setting is 4 hours
OFF = Disabled

TEMP CAL

On = Enabled. If set to On, set the
temperature variation from 0.2 to 4°C
OFF = Disabled

INT CAL
(Only on balances with
internal calibration masses)

YES = Use Internal calibration mass
NO = Use external calibration mass


Press [Mode] to advance to setting of the next menu which is
“PASSCODES” or [Esc] to return to normal weighing.

© Adam Equipment Company 2007