3 check weighing, Check weighing – Nova-Tech PGW Series User Manual
Page 18
Page 18 of 44
9.4.3 Check Weighing
During weighing of a sample the balance can be set to show if the weight is
above or below an upper and a lower limit. The display will use the arrows
under the fill guide to show the check weighing is operating. The arrows and
bars between the arrows will indicate when the weight is below the lower limit,
between the limits or above the upper limit.
The buzzer can be set to be active when the weight is outside the limits
(below the lower or above the upper) or within the limits (above the lower and
below the upper limit), or turned off. If desired, only one limit needs to be set.
If only one limit is set the other limit is considered to be zero (lower) or the
maximum (upper).
The Check weighing is not active from zero to 50d however the arrows will
still be turned on below the fill guide and the weight will still be displayed.
Steps:
• Press [Mode] and then the [Up] or [Down] key to select Check
Weighing. When “CHECK” is displayed press the [Enter] key to enter
Check Weighing mode
• Press [Up] or [Down] to set the “LOW LIMIT” to “On” or “OFF”
• Press [Enter] to proceed. If the “LOW LIMIT” was set to “On” the
display will show the current low limit and allow you to change the limit
using the numeric entry method (see section 7.1)
• Press [Enter] to proceed
• If the “LOW LIMIT” was set to “OFF” or the setting of the low limit is
complete, then the display will change to “HIGH LIMIT”. Use [Up] and
[Down] to set the “HIGH LIMIT” to “On” or “OFF”
• Press [Enter] to proceed
• If the “HIGH LIMIT” was set to “On”, the display will show the current
high limit which can be changed by using the numeric entry method
(see section 7.1)
• Press [Enter] to proceed
© Adam Equipment Company 2007