9.4.2 Percent Weighing
Percent weighing will be done by defining a certain weight to be 100%. The
weight to be used can either be entered by the user or taken from a sample
• Press [Mode] and then the [Up] or [Down] key to select Percent
weighing, “PERCENT” will be displayed
• Press [Enter] to enter the function
• Display will show, “PERCENT SAMPLE”
• Press [Enter] to select the sample method
• When “LOAD 100 %“ is shown, add the sample
• Press [Enter] to set this weight to be 100%, the busy symbol will turn
on. When ready the display will show “100%”
• Remove the sample and place an unknown sample to display the
percentage weight
• To set another weight as 100%, press [Mode] and continue as before
• To manually enter a value to be set as 100%, press [Up] or [Down] key
when “PERCENT SAMPLE” is shown to select “PERCENT Ent Wt”
• Press [Enter] to select the manually entered weight method
• Enter the weight using the numeric entry method (see section 7.1)
• Place unknown sample to display the percentage weight
• To perform percent weighing with another sample press [Mode] and
continue as before
• To return to normal weighing, press [Esc]
Note: Percentage will be displayed to the maximum number of decimal
places based on the resolution of the balance. To increase or decrease by
one decimal place, press the [Up] or [Down] key respectively.