13.0 SUPERVISOR MENUS
Pressing the [Setup] key while in normal weighing gives access to the
menus.
• When [Setup] is pressed and passcodes are not enable the display will
show the message “SUPERVISOR”. If passcodes are enabled, the
balance will ask for it by displaying “PASSCODE 0”
• If a wrong code is entered an “ERROR CODE” message will flash and
the balance will return to weighing mode
• If the passcode has been enabled and entered, the balance will allow
the operator to access the Supervisor’s menus
• When the display shows “SUPERVISOR” press [Enter] to view the
topics that can be modified
• From this menu the user can enable/disable weighing units or modes,
set balance parameters for the conditions, set time and date, set
parameters for the RS-232 interface, calibration parameters and
security parameters
•
The [Up] and [Down] keys will cycle through the main headings,
pressing the [Enter] key will enter the heading and sub-parameters or
options can be set. Press [Mode]
to come out of a sub-menu or to
[Esc] to return to normal weighing from any menu
13.1 ENABLE WEIGHING UNITS
• When “UNITS” is displayed, press [Enter]. The display will show the
symbol for the first unit, e.g. carats, ct, together with its enable state
“OFF” or “On”. The user can then enable or disable the carats unit by
using [Up] or [Down]. Pressing [Enter] will confirm the setting and will
advance to the next weighing unit. Repeat for each weighing unit in
turn. Gram is always set to “On”
• Press [Mode] to advance to setting of the next menu or press [Esc] to
return to normal weighing
© Adam Equipment Company 2007