10.2 AUTOMATIC CALIBRATION
The balance will have the ability to calibrate (or ask for calibration) when the 
balance has automatic calibration enabled and the conditions of the 
automatic calibration have been meet. 
 
Conditions that will cause an automatic calibration are: 
 
1. Internal temperature change greater than a preset amount, typically 2ºC 
2. Time since last calibration exceeds a preset time, typically 4 hours or 15 
minutes after power is applied.
 
On the balances with internal calibration the calibration will be done 
automatically when the balance is at a stable zero. . The user knows a 
calibration is asked for by the flashing of the “CAL” symbol on the display. 
The display will show a 5 second countdown when calibration will start. If the 
user presses the [Esc] key the calibration will be delayed one minute to allow 
time for a process to be finished. 
 
On balances with external calibration the balance will call for calibration to be 
carried out by flashing the “CAL” symbol on the display. As soon as the 
balance is calibrated the symbol will be turned off. 
 
The Auto calibration feature can be enabled, disabled or changed within the 
user options to meet the requirements of the users. 
 
 
10.3 CALIBRATION ERRORS
Occasionally during calibration an error will be detected. These errors can be 
caused by: 
 
Unstable readings 
Improper calibration weights being used 
Large shifts of zero from the factory settings 
 
When an error is found a displayed message will be shown and the 
calibration must be done again. If the balance has error messages more than 
once it is possible the mechanics have been damaged. 
 