13.4 SETUP PARAMETERS

The user parameters that control the balance are shown under the SETUP.
When “SETUP” is displayed, press the [Setup]/Enter key. The options for
each parameter can be scrolled through by using the [Up] or [Down] key.

LANGUAGE

English
French
German
Spanish

TIME

Set Time using the numeric entry
method (see section 7.1)

DATE DATE FORM
EUROPE (dd/mm/yy)
USA (mm/dd/yy)

Set Date using the numeric entry
method (see section 7.1)

INST ID

Enter a user number to identify this
balance

BUZZER

On= Enable
OFF= Disable

BACKLIGHT

On
OFF
AUTO

POWER DOWN

Set the time after which the unit will go
into Stand-by power settings,
On=Enable, OFF=Disable, If set to On-
the options are 1 to 9 minutes

FILTER

Set a value for the amount of filtering to
be done, set a value of 5, 10, 20, 30,
40 or 50. A larger number is more
filtering and a slower response. Default
is 20

STABILITY

Set a value to be used to determine
balance stability, set a value of 1, 2, 5
or 10d. A larger number corresponds to
a larger stable zone. Default is 5

AUTO ZERO

Can be set to On or OFF to enable the
auto-zero function. If set to On- select
from 1, 2 or 5d