13.2 ENABLE WEIGHING MODES
Same steps are followed to enable or disable the weighing modes.
• Press [Enter] when “MODES” is displayed. The display will show the
first mode i.e., Parts Counting (“PARTS”) together with its enabled state
“OFF” or “On”. The user can enable or disable the parts counting mode
by using [Up] or [Down]. Pressing [Enter] will confirm the setting and
will advance to the next weighing mode. Repeat for each mode in turn
• Press [Mode] to advance to setting of the next menu or press [Esc] to
return to normal weighing
13.3 ENABLE SERIAL INTERFACE PARAMETERS
The parameters affecting the serial setup are set in a similar manner to the
other parameters.
Press [Enter] when “SERIAL” is displayed to enter the sub-menu.
The parameters that can be set are:
ENABLE
The serial port can be set to On or OFF
BAUD RATE
Set the Baud Rate to 4800, 9600, 19200 or 38400
PARITY
Set the Parity to NONE, EVEN or ODD
STABLE
To print when stable (On) or regardless of stability (OFF)
CONTINUOUS Set the RS-232 to send data continuously to On or OFF
PERIODIC
Set the RS-232 to send data periodically (set in seconds)
to On or OFF. If On is selected, the value can be changed
between 1 and 999 seconds, using [Up] and [Down]
FORMAT
To send data as a single line of data, using the standard
format or using a customer designed format (FORM 1 or
FORM 2).