A typical test report will look as:
MOISTURE TEST 
FILE NAME: RES00032 
DATE: 13/12/2008 
TIME: 09:38:07 
SER NO: AE12345678 
 
TEST NO: _ _ _ _ _ _ _ _ _ 
 
USER NO: 9876543210 
 
Preset: PST 05 
 Bread-4 
 
Result: % Moisture 
Heating: Step 
Temp: Step 1 120 C 
 Time 1 04:00 
 Step 2 110 C 
 Time 2 90:00 
Rapid On 
Interval: 30 Sec 
Stop: Time/Stable 
 60 Min 
 0.002 g 
 30 Sec 
Start: Manual 
 
INIT MASS: 12.341 g 
 
MODE TEMP TIME RESULT 
Step0 65C 00:30 1.26 %M 
Step0 159C 01:00 2.11 %M 
Step0 157C 01:30 3.15 %M 
Step0 158C 02:00 3.79 %M 
Step0 156C 02:30 4.11 %M 
Step1 157C 03:00 4.19 %M 
Step1 149C 03:30 4.22 %M 
Step1 140C 04:00 4.24 %M 
Step2 131C 04:30 4.26 %M 
Step2 120C 05:00 4.27 %M 
Step2 111C 05:30 4.27 %M 
*********** AUTO STOP *********** 
 
LAST TEMP: 111C 
TEST TIME: 05:30 Min 
FINAL MASS: 11.820 g 
MASS LOSS: 0.521 g 
RESULT: 4.27 %M 
 
 
Heading 
File name used by analyzer to store result 
From Analyzer 
 
Serial number 
 
 
A blank line for users to fill in a number if they wish. 
 
User number 
 
Preset method if used. 
 
 
Test settings 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
Measured initial mass. 
 
Test Heading 
Results printed once every 30 seconds as set by the 
INCR time. Note temperature during first part of test is 
120C X 130% as the Rapid function was enabled. 
Step0 signifies the rapid function is active. After 3 
minutes the Step1 value is the target. 
 
 
 
 
 
 
 
TIME STOP if timed out or MAN STOP if user stopped 
 
Summary of the test 
 
 
 
 
 
English language version shown in the example. 
 