11.0 SERIAL
INTERFACE
The analyzers have the ability to send or receive data over the serial 
interface. Using either the RS-232 or the USB serial interface. 
The weighing data can be sent over the interface either automatically or when 
the user presses the [Print] key. 
The user has control over what data is to be printed.
11.1 RS-232 HARDWARE
The RS-232 interface is a simple 3 wire connection. The input and output 
connections are: 
Connector:
9 pin D-sub miniature plug
Pin
2
Input to analyzer
RXD
Pin
3
Output from analyzer TXD
Pin
5
Signal ground
GND
Handshaking is not applied.
Baud rate:
2400, 4800, 9600, 19200, 38400
Parity:
NONE (=8N1), EVEN (=8E1) or ODD (=8 O 1)
All lines are terminated with carriage return and line feed (
In continuous output mode, the serial output format will be a single line in the 
form “12.567 g
11.2 USB SERIAL HARDWARE
The USB interface is a standard USB connection. To use the USB interface it 
is necessary to plug the analyzer into the USB port of a PC. 
The PC will recognise the new device and search for a suitable driver. If 
none are found a message will be displayed to guide the user through the 
installation of suitable software. 