Lockout code prompt, Description, View-only status – Meriam M2110P SMART PRESSURE GAUGE User Manual
Page 22
22
Press the Up
↑ or Down↓ keys repeatedly to scroll
to the desired register value.
5
The Zero function (Up
↑ and Down↓) will reset the scroll to
default and allow editing to continue.
All available choices are scrolled.
(Choices indicate actual Deadband in percent of
full scale.)
6
Press the PRGM/Enter
→ key to accept the desired
value.
The value is accepted, and the register is closed.
Display shows “P8”.
7
Press the Backspace
← key to activate the register
setting and return to Measure Mode.
Normal Pressure Display.
Notes
1. After steps 2 and 3, the Backspace
← key will abort the process and return to Measure Mode.
2. After steps 4 and 5, the Backspace
← key will abort the edit and close the register.
3. After step 6, the PRGM/Enter
→ key will again open the register for edit.
4. During any programming operation, if there is no keypad activity for approximately 1 minute, the
operation is aborted and the gauge returns to Measure Mode unchanged.
15. Lockout Code Prompt
DESCRIPTION
When a gauge is “locked” for security purposes, it does not allow access to change its operating state
without first providing the lockout code. A gauge is “locked” by entering a lockout code in register P0
(See “STEP BY STEP: P0, LOCKOUT CODE” on page 11).
After pressing a key on a locked gauge for a desired function (ZERO, ENG UNITS, PRGM), “L 00”
will appear on the display, with the first 0 flashing. This is the prompt to enter the lockout code
(similar to a password).
VIEW-ONLY STATUS
If a user attempts to enter Program Mode without providing a correct lockout code (password), the
gauge briefly provides an error message, and then enters a “view-only” status, denoted by the PRGM
indicator flashing. In this mode, all registers (except the lockout code register itself) can be viewed,
but not changed.
In this view-only status, the operator navigates through the registers exactly the same as if
programming the registers. However, the keys that would change a value are simply ignored by the
gauge. Thus, the PRGM/Enter
→ key will open a register normally; the Backspace← key will close
the register, and then return to Measure Mode as expected.
STEP BY STEP: ENTERING A LOCKOUT CODE (When Prompted)
Step Action
Display
1
After attempting to re-zero, change Engineering
Units, or enter Program Mode...
The display will prompt “L 00” with the first zero
flashing for edit (indicating Lockout Code 00).
2
Press the Up
↑ or Down↓ keys repeatedly to scroll
to the desired numeric value for the flashing digit.
“L x0”, where “x” is the digit being edited.