User’s Manual Dwg No 9R141-IR

STEP BY STEP: P3, DENSITY OF WATER REFERENCE TEMP

Step Action

Display

1

Gauge should be in the normal Measure Mode. Normal

Pressure

Display.

Press the PRGM/Enter

→ key.

2

If the lockout is active, the gauge will now prompt for entry of
the lockout code. Refer to “STEP BY STEP: ENTERING A
LOCKOUT CODE (When Prompted)” on page 22.

The PRGM annunciator shows at the top of the
display, and the display shows the register name
P0”.

Press the Up

↑ or Down↓ keys repeatedly to scroll

to the desired register.

3

The Zero function (Up

and Down↓) will reset the scroll to P0

and allow selection to continue.

Display shows “P3”.

4

Press PRGM/Enter

→ to open the register.

The display shows the current value.

Press the Up

↑ or Down↓ keys repeatedly to scroll

to the desired register value.

5

The Zero function (Up

and Down↓) will reset the scroll to

default and allow editing to continue.

All available choices are scrolled.
“20” = 20 Deg. C
“4”

= 4 Deg. C

“60” = 60 Deg. F

6

Press the PRGM/Enter

→ key to accept the desired

value.

The value is accepted, and the register is closed.
Display shows “P3”.

7

Press the Backspace

← key to activate the register

setting and return to Measure Mode.

Normal Pressure Display.

Notes

1. After steps 2 and 3, the Backspace

← key will abort the process and return to Measure Mode.

2. After steps 4 and 5, the Backspace

← key will abort the edit and close the register.

3. After step 6, the PRGM/Enter

→ key will again open the register for edit.

4. During any programming operation, if there is no keypad activity for approximately 1 minute, the

operation is aborted and the gauge returns to Measure Mode unchanged.

11. P4 – User Scale All Models

The user scale feature is provided in addition to the standard Engineering Unit choices, to allow the display of
measured pressure in any Engineering Unit. The full scale value of the sensor is linearly re-scaled to the value
entered in this register. This is particularly useful for displaying linear tank levels in volumetric or mass units.

For example, a 20 PSIG gauge is programmed to read out in feet of water head by entering 46.22 in the P4
register. This sets the gauge to read from 0 to 46.22 feet of water, when User Units are selected as the
Engineering Units (46.22 feet of water at 20°C is equal to 20 PSIG).

User Engineering Units are selected when ALL Engineering Units are flashing during the Engineering Unit
selection process. Once selected, in Measure Mode, all Engineering Units are cleared from the display to
denote user-units.