beautypg.com

Step by step: p2, damp rate – Meriam M2110P SMART PRESSURE GAUGE User Manual

Page 14

background image

14

STEP BY STEP: P2, DAMP RATE

Step Action

Display

1

Gauge should be in the normal Measure Mode. Normal

Pressure

Display.

Press the PRGM/Enter

→ key.

2

If the lockout is active, the gauge will now prompt for entry of
the lockout code. Refer to “STEP BY STEP: ENTERING A
LOCKOUT CODE (When Prompted)” on page 22.

The PRGM annunciator shows at the top of the
display, and the display shows the register name
P0”.

Press the Up

↑ or Down↓ keys repeatedly to scroll

to the desired register.

3

The Zero function (Up

and Down↓) will reset the scroll to P0

and allow selection to continue.

Display shows “P2”.

4

Press PRGM/Enter

→ to open the register.

The display shows the current value.

Press the Up

↑ or Down↓ keys repeatedly to scroll

to the desired register value.

5

The Zero function (Up

and Down↓) will reset the scroll to

default and allow editing to continue.

All available choices are scrolled.

(Choices indicate actual damp rate in seconds.)

6

Press the PRGM/Enter

→ key to accept the desired

value.

The value is accepted, and the register is closed.
Display shows “P2”.

7

Press the Backspace

← key to activate the register

setting and return to Measure Mode.

Normal Pressure Display.

Notes

1. After steps 2 and 3, the Backspace

← key will abort the process and return to Measure Mode.

2. After steps 4 and 5, the Backspace

← key will abort the edit and close the register.

3. After step 6, the PRGM/Enter

→ key will again open the register for edit.

4. During any programming operation, if there is no keypad activity for approximately 1 minute, the

operation is aborted and the gauge returns to Measure Mode unchanged.

10. P3 – Density of Water Reference Temperature All Models

The M2110P supports pressure units of inches of water column and centimeters of water column, at three
different water temperature references: 20°C, 4°F and 60°F. The base reference temperature is 20°C on all
gauges. This register is used to change to one of the other standards. During programming, the corresponding
choices found in the register are “20”, “4”, and “60” respectively (the temperature unit cannot be shown due to
display limitations).


QUICK REFERENCE: P3, DENSITY OF WATER REFERENCE TEMP

Px

Name

Description

Value Range

Notes

P3

Density of Water
Reference
Temperature

Sets reference temperature for
water density when gauge is set
for water column units.

20°C
4°C
60°F