Step by step: p5, setpoint options – Meriam M2110P SMART PRESSURE GAUGE User Manual
Page 18
18
STEP BY STEP: P5, SETPOINT OPTIONS
Step Action
Display
1
Gauge should be in the normal Measure Mode. Normal
Pressure
Display.
Press the PRGM/Enter
→ key.
2
If the lockout is active, the gauge will now prompt for entry of
the lockout code. Refer to “STEP BY STEP: ENTERING A
LOCKOUT CODE (When Prompted)” on page 22.
The PRGM annunciator shows at the top of the
display, and the display shows the register name
“P0”.
Press the Up
↑ or Down↓ keys repeatedly to scroll
to the desired register.
3
The Zero function (Up
↑ and Down↓) will reset the scroll to P0
and allow selection to continue.
Display shows “P5”.
4
Press PRGM/Enter
→ to open the register.
The display shows the current value.
Press the Up
↑ or Down↓ keys repeatedly to scroll
to the desired register value.
5
The Zero function (Up
↑ and Down↓) will reset the scroll to
default and allow editing to continue.
All available choices are scrolled.
6
Press the PRGM/Enter
→ key to accept the value.
The value is accepted, and the register is closed.
Display shows “P5”.
7
Press the Backspace
← key to activate the register
setting and return to Measure Mode.
Normal Pressure Display.
Notes
1. After steps 2 and 3, the Backspace
← key will abort the process and return to Measure Mode.
2. After steps 4 and 5, the Backspace
← key will abort the edit and close the register.
3. After step 6, the PRGM/Enter
→ key will again open the register for edit.
4. During any programming operation, if there is no keypad activity for approximately 1 minute, the
operation is aborted and the gauge returns to Measure Mode unchanged.