beautypg.com

Quick reference: p6/p7, setpoints, Step by step: p6/p7, setpoints – Meriam M2110P SMART PRESSURE GAUGE User Manual

Page 20

background image

20

QUICK REFERENCE: P6/P7, SETPOINTS

Px

Name

Description

Value Range

Notes

SetPoint model:

Controls

SET1

relay.

P6 SET1

Current Loop model:

Sets 4.00 mA value.

–20% to +120%
of Full Scale

SetPoint model:

Controls

SET2

relay.

P7 SET2

Current Loop model:

Sets 20.00 mA value.

–20% to +120%
of Full Scale

User defined
values.

On Current
Loop models,
SET1 cannot
equal SET2.

STEP BY STEP: P6/P7, SETPOINTS

Step Action

Display

1

Gauge should be in the normal Measure Mode. Normal

Pressure

Display.

Press the PRGM/Enter

→ key.

2

If the lockout is active, the gauge will now prompt for entry of
the lockout code. Refer to “STEP BY STEP: ENTERING A
LOCKOUT CODE (When Prompted)” on page 22.

The PRGM annunciator shows at the top of the
display, and the display shows the register name
P0”.

Press the Up

↑ or Down↓ keys repeatedly to scroll

to the desired register.

3

The Zero function (Up

and Down↓) will reset the scroll to P0

and allow selection to continue.

Display shows “P6” or “P7”.

Also, “SET1” or “SET2” respectively is
displayed to help identify the register, in addition
to “4-20mA” for Current Loop Models.

4

Press PRGM/Enter

→ to open the register.

The display will show the current value, with the
first zero flashing for edit.

Press the Up

↑ or Down↓ keys repeatedly to scroll

to the desired numeric value for the flashing digit.

5

The Zero function (Up

and Down↓) will reset the entire

register to 0000 and allow editing to continue. The decimal
point and negative sign will not be affected, and same digit will
be flashing.

_xxx”, where “_” is the digit being edited.

6

Press the PRGM/Enter

→ key to accept the digit.

The next digit begins flashing for edit.

Repeat steps 5 and 6 for all the remaining digits.

7

The Backspace

← key will backup one digit at a time, or abort

and close the register if the first digit is flashing.

The PRGM/Enter

→ key will accept one digit at a time, moving

to the next, or accept the complete value and close the register
after the decimal point edit.

After Enter

→ when the last digit is flashing, the

decimal point flashes for edit.

Please read carefully note 4 on page 17.