P8 – deadband set point models only, Quick reference: p8, deadband, Step by step: p8, deadband – Meriam M2110P SMART PRESSURE GAUGE User Manual
Page 21
21
User’s Manual Dwg No 9R141-IR
Press the Up
↑ or Down↓ keys repeatedly to scroll
to the desired decimal point position, and to select
positive/negative.
8
If the Zero function (Up
↑ and Down↓) is used during decimal
point editing, it will still reset the register’s value to 0000, and
not affect the decimal point position or negative sign.
Decimal point moves accordingly.
Please read carefully note 4 on page 17.
Negative Sign activates and de-activates when the
decimal point sequence is scrolled through
completely.
9
Press the PRGM/Enter
→ key to accept the
complete value.
The value is accepted, and the register is closed.
Display shows “P6” or “P7”.
10
Press the Backspace
← key to activate the register
setting and return to Measure Mode.
Normal Pressure Display.
Notes
Please refer to the “notes” section for P4 (User Scale) on page 17. The editing process is the same.
14. P8 – DEADBAND Set Point Models Only
This register is found on the SetPoint model only. It sets the adjustable deadband for the SPDT relays.
Deadband can be set to 0%, 0.1, 0.2, 0.5, 1, 2, 5 and 10% of full scale. The relay will energize precisely at the
value in its corresponding P6 and P7 registers on increasing pressure. The relay will de-energize at a value
equal to the corresponding P6 and P7 register, minus the deadband value, on decreasing pressure. With a 0%
deadband setting, the relays will energize and reset precisely at the values in the corresponding P6 and P7
registers. During programming, the choices found in the register correspond to the actual deadband values in
percent full scale.
QUICK REFERENCE: P8, DEADBAND
Px
Name
Description
Value Range
Notes
P8
Deadband
Sets the amount of deadband in
percent of full scale for relays.
0, 0.1, 0.2, 0.5, 1,
2, 5, 10% FS
SetPoint model only.
0 = Disabled.
STEP BY STEP: P8, DEADBAND
Step Action
Display
1
Gauge should be in the normal Measure Mode. Normal
Pressure
Display.
Press the PRGM/Enter
→ key.
2
If the lockout is active, the gauge will now prompt for entry of
the lockout code. Refer to “STEP BY STEP: ENTERING A
LOCKOUT CODE (When Prompted)” on page 22.
The PRGM annunciator shows at the top of the
display, and the display shows the register name
“P0”.
Press the Up
↑ or Down↓ keys repeatedly to scroll
to the desired register.
3
The Zero function (Up
↑ and Down↓) will reset the scroll to P0
and allow selection to continue.
Display shows “P8”.
4
Press PRGM/Enter
→ to open the register.
The display shows the current value.