Quick reference: p4, user scale, Step by step: p4, user scale – Meriam M2110P SMART PRESSURE GAUGE User Manual
Page 16
16
QUICK REFERENCE: P4, USER SCALE
Px
Name
Description
Value Range
Notes
P4 User
Units
Full Scale
Enables gauge to display units
other than factory defined
Engineering Units.
0 to 9,999 (Keypad)
0 to 19,999 (Serial)
User defined
value.
STEP BY STEP: P4, USER SCALE
Step Action
Display
1
Gauge should be in the normal Measure Mode. Normal
Pressure
Display.
Press the PRGM/Enter
→ key.
2
If the lockout is active, the gauge will now prompt for entry of
the lockout code. Refer to “STEP BY STEP: ENTERING A
LOCKOUT CODE (When Prompted)” on page 22.
The PRGM annunciator shows at the top of the
display, and the display shows the register name
“P0”.
Press the Up
↑ or Down↓ keys repeatedly to scroll
to the desired register.
3
The Zero function (Up
↑ and Down↓) will reset the scroll to P0
and allow selection to continue.
Display shows “P4”.
4
Press PRGM/Enter
→ to open the register.
The display will show the current value, with the
first digit flashing for edit.
Press the Up
↑ or Down↓ keys repeatedly to scroll
to the desired numeric value for the flashing digit.
5
The Zero function (Up
↑ and Down↓) will reset the entire
register to 0000 and allow editing to continue. The decimal
point will not be affected, and same digit will be flashing.
“_xxx”, where “_” is the digit being edited
(flashing).
6
Press the PRGM/Enter
→ key to accept the digit.
The next digit begins flashing for edit.
Repeat steps 5 and 6 for the all remaining digits.
7
The Backspace
← key will backup one digit at a time, or abort
and close the register if the first digit is flashing.
The PRGM/Enter
→ key will accept one digit at a time, moving
to the next, or accept the complete value and close the register
after the decimal point edit.
After Enter
→ when the last digit is flashing, the
decimal point flashes for edit.
Please read carefully note 4 below.
Press the Up
↑ or Down↓ keys repeatedly to scroll
to the desired decimal point position.
8
If the Zero function (Up
↑ and Down↓) is used during decimal
point editing, it will still reset the register’s value to 0000, and
not affect the decimal point position.
Decimal point moves accordingly.
Please read carefully note 4 below.
9
Press the PRGM/Enter
→ key to accept the
complete value.
The value is accepted, and the register is closed.
Display shows “P4”.
10
Press the Backspace
← key to activate the register
setting and return to Measure Mode.
Normal Pressure Display.