P0 – lockout code all models, Quick reference: p0, lockout code, Step by step: p0, lockout code – Meriam M2110P SMART PRESSURE GAUGE User Manual
Page 11
11
User’s Manual Dwg No 9R141-IR
3
Press PRGM/Enter
→ before the countdown
expires
1
. This acknowledges the warning.
The display will again countdown from
“0005” to “0000”.
4
Press ZERO again (Up
↑ and Down↓) before the
countdown expires
1
. This confirms the zero action.
The display will flash “0000” several times while
the new zero is taken.
Notes
1. During steps 2 or 3, if the countdown expires, the gauge will return to Measure Mode unchanged.
2. During steps 2 or 3, the Backspace
← key will abort the process and return to Measure Mode.
7. P0 – Lockout Code All Models
This feature provides security in the M2110P. It is designed to prevent unauthorized personnel from tampering
with or inadvertently changing the configuration of the gauge. The lockout is controlled by a 2-digit setting in
the P0 register. Once a lockout code is entered, the gauge will prompt for the lockout code before allowing any
changes (similar to a “password”). Changes include Engineering Unit selection, re-zeroing, or entering
Program Mode.
Following an operator action, if the correct code (password) is not entered when prompted, an error message is
briefly displayed. If the operator action was changing Engineering Units or re-zero the gauge, the gauge will
simply return to normal Measure Mode operation, without accepting any change. If the operator action was
entering Program Mode, the gauge will enter a “view-only” status (see page 22), denoted by the PRGM
indicator flashing.
In Program Mode (and during lockout code prompting), the register value is shown as “L xx” to assist in
identifying the register. The “L ” indicates “Lockout”, and “xx” will consist of the current value.
QUICK REFERENCE: P0, LOCKOUT CODE
Px
Name
Description
Value Range
Notes
P0
Lockout Code
Lockout for security.
00 to 99
00 = Disabled.
STEP BY STEP: P0, LOCKOUT CODE
Step Action
Display
1
Gauge should be in the normal Measure Mode. Normal
Pressure
Display.
Press the PRGM/Enter
→ key.
2
If the lockout is active, the gauge will now prompt for entry of
the lockout code. Refer to “STEP BY STEP: ENTERING A
LOCKOUT CODE (When Prompted)” on page 22.
The PRGM annunciator shows at the top of the
display, and the display shows the register name
“P0”.
3
Press PRGM/Enter
→ again, to open the register.
The display will show “L xy” with the first digit
flashing for edit (indicating Lockout Code xy).
4
Press the Up
↑ or Down↓ keys repeatedly to scroll
to the desired numeric value for the flashing digit.
“L _x”, where “_” is the digit being edited
(flashing).