Revive after storage routine – Luminex MAGPIX User Quick Guide User Manual
Page 9
FIGURE 1.
Sample Probe Height Adjustment
Revive After Storage Routine
NOTE: The Revive After Storage routine is necessary when the system
runs for the first time and is recommended when the system has
been idle for more than a week.
After you have adjusted the sample probe height, run the Revive After Storage (Luminex)
routine.
1. Open the Maintenance page, then the Cmds & Routines tab.
2. Select Revive After Storage (Luminex) from the drop-down list. The Revive After
Storage routine performs the following commands:
• Prime
• Rinse
• Alcohol Flush
• Rinse
3. Add 70% isopropanol or ethanol to reservoir RB1 on the off-plate reagent block as
indicated on the Cmds & Routines tab.
NOTE: The rinse reservoir (RD1) should be empty.
4. Click Retract.
5. Click Run.
After the Revive After Storage routine is complete, run the System Initialization routine.
MAGPIX
®
User Quick Guide 4.2
4