3. Under Maintenance Options, select Fluidics prep, performance verification from the

drop-down list.

FIGURE 2.

Setting System Initialization Routine

4. Click Save.

Running the System Initialization Routine

1. On the Home page, click System Initialization under Daily Activities. The Auto Maint

tab opens. On the Auto Maint tab, the System Initialization option is automatically
selected.

2. Verify that the correct lot kit is displayed in the Calibration and Performance

Verification drop down field and that the correct reagents (for example, VER and
Fluidics reagents) have been added to the off-plate reagent block.

3. Fill reservoir RB1 3/4 full of 70% isopropanol or ethanol.

4. Verify that the Rinse reservoir RD1 is empty.

5. Click Retract.

6. Click Run.