MAGPIX New System Startup Errors

Problem

Resolution

1. Check the sample probe height for correctness.

Calibration Failure

2. Create and run the following routine.

Create routines in the xPONENT Maintenance page Cmds & Routines tab.

Approximate run time: Five minutes

(2x) Clean with 70% isopropanol or ethanol alcohol (reservoir B)

(2x) Alcohol Flush with 70% isopropanol or ethanol alcohol (reservoir B)

(3x) Rinse (reservoir D)

(3x) Prime

Run Calibration routine

Verify that reagent is freshly vortexed. Use 6 drops of reagent.

3. If calibration continues to fail, create and run the following routine.

Create routines in the xPONENT Maintenance page Cmds & Routines tab.

Approximate run time: Five minutes

MAGPIX

®

User Quick Guide 4.2

18