Magpix new system startup errors, Magpix off-plate reagent block – Luminex MAGPIX User Quick Guide User Manual
Page 23
FIGURE 3.
MAGPIX Off-Plate Reagent Block
1
Reservoirs
2
Strip Wells
MAGPIX New System Startup Errors
Problem
Resolution
1. Check the sample probe height for correctness.
Calibration Failure
2. Create and run the following routine.
Create routines in the xPONENT Maintenance page Cmds & Routines tab.
Approximate run time: Five minutes
(2x) Clean with 70% isopropanol or ethanol alcohol (reservoir B)
(2x) Alcohol Flush with 70% isopropanol or ethanol alcohol (reservoir B)
(3x) Rinse (reservoir D)
(3x) Prime
Run Calibration routine
Verify that reagent is freshly vortexed. Use 6 drops of reagent.
3. If calibration continues to fail, create and run the following routine.
Create routines in the xPONENT Maintenance page Cmds & Routines tab.
Approximate run time: Five minutes