Laurel Electronics Custom ASCII Protocol For Laureate Series 2 Digital Panel Meters, Counters, Timers & L-Series Transmitters User Manual

Page 40

background image

40

3D (L)

Analog Setup

Bit 7 6 5 4 3 2 1 0 Analog Output Source

0 0 Filtered Item

0 1 Item 1

1 0 Item 2

1 1 Item 3

0

0 to 20mA output

1

0 to 10V output

1 0

4 to 20mA output

1 1

-10V to 10V output

3C (L)

Source

Bit 7 6 5 4 3 2 1 0

Compare Setpoint 2 to:

0 0 Filtered Item

0 1 Item 1

1 0 Item 2

1 1 Item 3

Compare Setpoint 1 to:

0 0

Filtered Item

0 1

Item 1

1 0

Item 2

1 1

Item 3

36 (L)

Lockout2
0 = unlocked
1 = locked

Bit 7 6 5 4 3 2 1 0

1 Change Item #

1

CALib

1

Ser 1, Ser 2, Ser 3

1

An Lo, An Hi, An SEt

1

Front Panel meter reset

1

Front Panel Peak, Latched resets

1

View alarm setpoints

1

View Peak locked

35 (L)

Lockout1
0 = unlocked
1 = locked

Bit 7 6 5 4 3 2 1 0

1 FiLtEr

1

Gate t, time out, batch, pulses

1

SEtuP, ConFiG, display number

1

InPut

1

Change Setpoints

1

SourcE,AL SEt,dEVn1b,1H,2b,2H

1

SCALE, OFFSEt, Coords, rESoLn

1

SLOPE, dECPt