Laurel Electronics Custom ASCII Protocol For Laureate Series 2 Digital Panel Meters, Counters, Timers & L-Series Transmitters User Manual

Page 31

background image

31

69 (L)

Serial Cnfg3

Bit 7 6 5 4 3 2 1 0

0 0 0 Send Reading

0 0 1 Send Peak

0 1 0 Send Valley

0 1 1 Send Reading + Peak

1 0 0 Send Reading + Valley

1 0 1 Send Reading + Peak + Valley

0

or at end of all Items

1

or at end of each Item

(if no Alarm character)

0

Non-latching RTS

1

Latching RTS

0

Normal continuous TX

1

Special Start & Stop characters

0

Full duplex

1

Half duplex

35 (L)

Decimal Point

01 Byte values in hex

XXXXX.

02 (2 hex characters/byte) XXXX.X

03

XXX.XX

04

XX.XXX

05

X.XXXX

06

.XXXXX

34 (L)

Lockout2
0 = unlocked
1 = locked

Bit 7 6 5 4 3 2 1 0 Menu item & front panel lockout

1 Serial configuration

1

Analog output scaling

1

Alarm setpoint programming

1

Alarm setup

1

Front panel DPM reset

1

Front panel Peak & Alarm reset

1

View alarm setpoints

1

View Peak value & Tare function

33 (L)

Lockout1
0 = unlocked
1 = locked

Bit 7 6 5 4 3 2 1 0 Menu item & front panel lockout

1 Offset, Lo & Hi readings

1

Scale, Lo In, Hi In

1

Filter Setup

1

Setup, Config & Decimal Point

1

InPut Menu Item