Data inquiry, Character 1 - stx, Characters 2, 3 - device – CONTREX ML-TRIM User Manual
Page 96: Character 4 - message type
3 - 62
Data Inquiry
Use the Data Inquiry to request the current value for Parameters (i.e., Control
Parameters or Monitor Parameters).
Table 3-33 Data Inquiry - Host Transmission
Character #
1
2
3
4
5
6
7
8
9
10
11
12
DEV # DEV # MSG
PAR # PAR #
DATA DATA
DATA
DATA
DATA
DESC
STX
10s
1s
TYPE
10s
1s
1000s 100s
10s
1s
FORM
ETX
ASCII
STX
0-9
0-9
2
0-9
0-9
0
0
0
0
0
ETX
The following is a description of the Data Inquiry - Host Transmission Characters.
Character 1 - STX:
This is the first character in the character string. None of the other characters will
be recognized without this character prefix. Always use the ASCII “STX” character;
it enables the ML–Trims receive buffer.
Characters 2, 3 - Device #:
These characters are the access address of the ML–Trim. This number identifies
individual ML–Trims on a mutltidrop system. The ML–Trim will accept data only if
this number matches the ML–Trim's address (CP-70), with the exception of a “00”
address. The “00” address is universally accepted by all ML–Trims that are on the
RS485 Serial Communications Interface.
Character 4 - Message Type:
This character should always be “2”.