Basic user manual, Specifications, English – Hach-Lange ORBISPHERE 6110 Basic User Manual User Manual
Page 3

Specifications
Specifications are subject to change without notice
Specification
Details
Measurement range
O
2
1 ppb - 12 ppm
CO
2
0.75 - 5 V/V or 1.5 - 10 g/kg
Repeatability r
95
Total Package Oxygen
± 5 μg/L ± 10% whichever is the
greater
CO
2
on equilibrated packages at T
= 10 to 25°C (50 to 77°F)
± 0.05 V/V or 0.10 g/kg ± 2%
whichever is the greater
O
2
± 2 ppb ± 10% whichever is the
greater
Typical analysis time
About 4 minutes
Display units
O
2
concentration
ppb or ppm
CO
2
concentration
V/V, g/kg, g/L or %W
Pressure
bar, mbar, psia
Temperature
°C, °F or K
Operating limits
Package temperature
-2 to 30°C (28 to 86°F)
Package pressure
1.4 to 6.8 bar absolute (20 - 99 psia)
Ambient temperature
0 to 40°C (32 to 104°F)
Relative humidity
Up to 80%
Package setup
Maximum package height
340 mm (13.39 ins)
Minimum package height
90 mm (3.54 ins)
Minimum volume
150 ml
Material
Glass, PET or aluminum
Dimensions (L x W x H)
537 x 540 x 942 mm (21.1 x 21.3 x 37.1 ins)
Weight
55 kg (121 lbs)
Enclosure protection
IP20
Power requirements
100-240 VAC ±10% @ 50-60 Hz
Power consumption
Max. 250 VA
Purge gas
CO
2
with purity > 99.9% at 6 to 7 bar absolute (87 to 102 psia)
Purge gas consumption
0.4 mL/second (1.5 L/hour)
Forcing gas
Air or N
2
at 5.5 to 6 bar absolute (80 to 87 psia)
Forcing gas consumption
0.25 mL/minute in standby mode
European directives
Low voltage 2006/95/EC
EMC 2004/108/EC
Electromagnetic compatibility
standards
EN 61326: 2006
Safety standard
IEC/UL/CSA 61010-1
Laser product safety
IEC/UL/CSA 60825-1
English 3