ST100 Series Flow Meter
INSTALLATION
Fluid Components International LLC
25
Alarm Activation/Deactivation – tag AL
When an alarm condition either activates (alarm condition met) or deactivates (alarm condition no longer present), an alarm log entry will be
generated.
Following the tag field, the data items are:
• Alarm Id (1..6)
• Activated (1) or Deactivated (0)
• Alarm mode:
o Flow high: 1
o Flow low: 2
o Temperature high: 3
o Temperature low: 4
o Pressure high: 5
o Pressure low: 6
• Alarm process data value
o If Alarm mode 1 or 2, then flow (in customer units)
o If Alarm mode 3 or 4, then temperature (in customer units)
o If Alarm mode 5 or 6, then pressure (in customer units)
year month
day
time
tag
Alarm
ID
Act/
De-Act
Mode PD value fault code
fault code
fault code
2011
5
24
13:44:09
AL
1
1
2
0
0x00100000 0x00000001 0x00000000
2011
5
24
13:44:39
AL
2
1
3
86.63182 0x00100000 0x00000001 0x00000000
Log File Entry Example: Alarm 1 and 2 activated
FE Self Test: Delta-R – tag DR
When an automatic Delta-R test for an FE is executed, the results are logged with a Delta-R log entry.
Following the tag field, the data items are:
• FE Id (0..15)
• Delta-R low measured flow value (in customer units)
• Delta-R low expected flow value (in customer units)
• Delta-R medium measured flow value (in customer units)
• Delta-R medium expected flow value (in customer units)
• Delta-R high measured flow value (in customer units)
• Delta-R high expected flow value (in customer units)
year month day
time
tag
FE ID DR low m DR low e DR med m DR low e
DR hi m
DR hi e
2011
5
24
13:44:39
DR
0
60.48953
60.48953
100.48953 100.48953 150.48953 150.48953
Log File Entry Example: Automatic Delta R Calibration Check
Note: This log function is setup in the Diagnostic menu under the Scheduled Tests tab
Two other tag codes will be logged if the instrument identifies a system fault. These additional tags are FL and CF. If these fault tags are
identified in a log file, contact FCI customer service.