Fluid Components International CMF Series Guide User Manual
Page 9
Doc. No. 06EN003318 Rev. B 9 FlexCOR
TM
Model CMF Series Coriolis Flow Meter
FLUID COMPONENTS, INTL
IMPORTANT: Access to the change mode is protected by a user code. The user code is factory set to 1000, but can
be changed to any value between 1000 and 9999 in the change password menu.
The factory setting of 1000 can be reestablished by switching off power . Then press the TOP UP key while switching
on the power. The user code is now reset to 1000.
Example of Programming of Max Mass Flow and Event Changes
Basic settings menu description:
This menu is used for basic configuration of the instrument with a choice of units, minimum and maximum limits for display
and analog/digital outputs for all measurement parameters, i.e. mass flow, volume flow, fraction, temperature and density.
Settings of min./max. values and units:
Numerical values are entered by placing the cursor in the field that is to be set using the SELECT key. Press unlock and the
value can be changed using the change key. The desired value is locked by activating LOCK.
Positioning of the decimal point is carried out by placing the cursor below the decimal point using the SELECT key. The position
can be set using the set key. The LOCK key is activated and the decimal point is now positioned.
Selecting the unit:
Place the cursor below the unit using SELECT key. Set the desired unit using CHANGE key. Press the LOCK key to save the
setting. Place the cursor below the time scale using SELECT key . Choose the desired time scale using CHANGE. Save the
value by pressing the LOCK key.
The maximum and minimum values set will then apply to all outputs, e.g. where the min. value will correspond to 0-4 mA
depending on the setting of the current output and the max. will correspond to 20 mA.
As example, to change the default setting of the maximum mass flow on a CMF-A from 20 Kg/h to .45 lb/min, do the following:
Keypad operation
Implementation
Display on Transmitter
Push for 2 sec.
To access the
Password
user password
0000
Push once
To unlock password
CHANGE
0000
Push once
To enter 1000 as
CHANGE
password
1000
Push once
To lock password and to
CONV.SETUP MODE>
enter the menu
Basic settings
Push once
To enter basic setting
Flow direction
submenu
Positive
Push once
To go to mass flow
Massflow max.
max. setting
000020. kg/h
Push once
To change num.
Massflow max.
value
000020. kg/h
Push 4 times
To move the cursor
Massflow max.
to the num. position
000020. kg/h
Push
Until 4 appears
Massflow max.
000040. kg/h
Push once
To move the cursor to
Massflow max.
the next num. position
000040. kg/h
Push
Till 5 appears
Massflow max.
000045. kg/h
Push once
To move the cursor to
Massflow max.
the decimal point
000045. kg/h
Push
To position the decimal
Massflow max.
point correct
0000.45 kg/h
Push 3 times
To move cursor
Massflow max.
to "Kg" unit
0000.45 kg/h
Push twice
To change units to lb.
Massflow max.
0000.45Lb/h
Push once
To move cursor to
Massflow max.
the "h" unit
0000.45Lb/h
Push 3 times
To change "h" to "min"
Massflow max.
0000.45 Lb/min
Push
To lock the new setting
Massflow max.
of the mass flowmeter
000.45 Lb/m
Push twice
Transmitter reverts to
standard operation
MENU CMDS