beautypg.com

Fluid Components International CMF Series Guide User Manual

Page 9

background image

Doc. No. 06EN003318 Rev. B 9 FlexCOR

TM

Model CMF Series Coriolis Flow Meter

FLUID COMPONENTS, INTL

IMPORTANT: Access to the change mode is protected by a user code. The user code is factory set to 1000, but can

be changed to any value between 1000 and 9999 in the change password menu.

The factory setting of 1000 can be reestablished by switching off power . Then press the TOP UP key while switching

on the power. The user code is now reset to 1000.

Example of Programming of Max Mass Flow and Event Changes

Basic settings menu description:

This menu is used for basic configuration of the instrument with a choice of units, minimum and maximum limits for display
and analog/digital outputs for all measurement parameters, i.e. mass flow, volume flow, fraction, temperature and density.

Settings of min./max. values and units:
Numerical values are entered by placing the cursor in the field that is to be set using the SELECT key. Press unlock and the

value can be changed using the change key. The desired value is locked by activating LOCK.

Positioning of the decimal point is carried out by placing the cursor below the decimal point using the SELECT key. The position

can be set using the set key. The LOCK key is activated and the decimal point is now positioned.

Selecting the unit:

Place the cursor below the unit using SELECT key. Set the desired unit using CHANGE key. Press the LOCK key to save the

setting. Place the cursor below the time scale using SELECT key . Choose the desired time scale using CHANGE. Save the
value by pressing the LOCK key.

The maximum and minimum values set will then apply to all outputs, e.g. where the min. value will correspond to 0-4 mA
depending on the setting of the current output and the max. will correspond to 20 mA.

As example, to change the default setting of the maximum mass flow on a CMF-A from 20 Kg/h to .45 lb/min, do the following:

Keypad operation

Implementation

Display on Transmitter

Push for 2 sec.

To access the

Password

user password

0000

Push once

To unlock password

CHANGE
0000

Push once

To enter 1000 as

CHANGE

password

1000

Push once

To lock password and to

CONV.SETUP MODE>

enter the menu

Basic settings

Push once

To enter basic setting

Flow direction

submenu

Positive

Push once

To go to mass flow

Massflow max.

max. setting

000020. kg/h

Push once

To change num.

Massflow max.

value

000020. kg/h

Push 4 times

To move the cursor

Massflow max.

to the num. position

000020. kg/h

Push

Until 4 appears

Massflow max.
000040. kg/h

Push once

To move the cursor to

Massflow max.

the next num. position

000040. kg/h

Push

Till 5 appears

Massflow max.
000045. kg/h

Push once

To move the cursor to

Massflow max.

the decimal point

000045. kg/h

Push

To position the decimal

Massflow max.

point correct

0000.45 kg/h

Push 3 times

To move cursor

Massflow max.

to "Kg" unit

0000.45 kg/h

Push twice

To change units to lb.

Massflow max.
0000.45Lb/h

Push once

To move cursor to

Massflow max.

the "h" unit

0000.45Lb/h

Push 3 times

To change "h" to "min"

Massflow max.
0000.45 Lb/min

Push

To lock the new setting

Massflow max.

of the mass flowmeter

000.45 Lb/m

Push twice

Transmitter reverts to
standard operation

MENU CMDS