Notice of proprietary rights, Error codes – Fluid Components International CMF Series Guide User Manual
Page 12
FLUID COMPONENTS, INTL
FlexCOR
TM
Model CMF Series Coriolis Flow Meter 12 Doc. No. 06EN003318 Rev. B
Notice of Proprietary Rights
This document contains confidential technical data, including trade secrets and proprietary information which are the property of Fluid
Components Intl (FCI). Disclosure of this data to you is expressly conditioned upon your assent that its use is limited to use within your company
only (and does not include manufacture or processing uses). Any other use is strictly prohibited without the prior written consent of FCI.
Visit FCI on the Worldwide Web: www.fluidcomponents.com
Error Error text
#Comment
Outputs
Input
No.
Remedy text
status
status
1
I1 - Power on
OK
Power on has activated.
Active
Active
2
I2 - Add-on Module
Applied
A new module has been added to the system.
Active
Active
3
I3 - Add-on Module
An add-on module is bad or has been removed.
Install
This can also be an internal add-on module.
Active
Active
4
I4 - Param. corrected
A less vital parameter in the converter has been
OK
replaced by its default value.
Active
Active
20
W20 - Totalizer 1
During initialization, the check of the saved totalize
Reset manually
value failed. The saved totalizer value is not reliable.
Reset the totalizer value manually to rely on readings.
Active
Active
20
W20 - Totalizer 2
During initialization, the check of the saved totalize
Reset manually
value failed. The saved totalizer value is not reliable.
Reset the totalizer value manually to rely on readings.
Active
Active
21
W21 - Pulse overflow
Actual flow is too big compared with pulse width and
Reduced
Adjust pulse settings
mass/pulse.
pulse width
Active
22
W22 - Batch timeout
Duration of batching has exceeded a predefined
Batch out-
Check installation
maximum time.
put on zero
Active
23
W23 - Batch overrun
Batch quantity has exceeded a predefined maximum
Batch out-
Check installation
overrun mass or volume.
put on zero
Active
24
W24 - Batch neg. flow
Check flow direction
Negative flow direction during batch.
Active
Active
30
W30 - Flowsaturated
Adjust max. flow
Flow is above Q
max.
settings.
Max. 120 %
Active
31
W31- Empty pipe
Pipe is empty.
Zero
Active
32
W32 - Temp. to high
The temperature of the fluid has exceeded the max.
Adjust temperature
temperature rating of the sensor (180
°C).
Active
Active
33
W33 - Temp. to low
The temperature of the fluid has exceeded the min.
Adjust temperature
temperature rating of the sensor (-50
°C).
Active
Active
34
W34 - Zero Adj. fail
Zero-point values are out of limits because of flow in the
Check flow = zero
sensor. Check zero flow conditions, valves, pumps etc.
Active
Active
35
W35 - Current Out 1
Current output exceeds 120%. Ensure that the sensor is
Check max. settings
correctly sized and check maximum flow setting.
Active
Active
36
W36 - Freq/Pulse Out1
Freq/Pulse output exceeds 120%. Ensure that the sensor
Check max. settings
is correctly sized and check maximum flow setting.
Active
Active
40
P40 - SENSORPROM
â
Insert
SENSORPROM
â
unit not installed.
Active
Active
41
P41 - Parameter range
A parameter is out of range.
Switch off and on
The error will disappear at the next power-on.
Active
Active
42
P42 - Current output
Current loop is disconnected or the loop resistance
Check cables
is too big.
Active
Active
43
P43 - Internal error
Internal error.
Switch off and on
Active
Active
49
P49 - Protec. viol.
Too many errors occured at the same time.
Switch off and on
Some errors are not detected correctly.
Active
Active
50
P50 - Temp. cable
Error in temperature sensor, check cables and
Check cable
connectors.
Active
Active
51
P51 - Pick-up 1
Pick-up 1 amplitude too low. Check cables or application
Check cable/install.
for damping (air/gas in liquid).
Active
Active
52
P52 - Pick-up 2
Pick-up 2 amplitude too low. Check cables or application
Check cable/install.
for damping (air/gas in liquid).
Active
Active
60
F60 - CAN comm. error
CAN bus communication error. An add-on module, the
Converter/add-on module
display module or the converter is defect
Zero
Inactive
61
F61 - SENSORPROM
â
err.
SENSORPROM
â
data is not reliable.
Replace
Active
Active
62
F62 - SENSORPROM
â
ID
SENSORPROM
â
unit ID doesn't match instrument ID.
Replace
The SENSORPROM
â
unit is from another instrument.
Zero
Inactive
63
F63 - SENSORPROM
â
It is not possible to read from the SENSORPROM
â
Replace
unit .
Active
Active
70
F70 - Pick-up phase
Check cables/polarity.
Active
Active
71
F71 - Driver phase
Check cables/polarity.
Active
Active
80-83
F80, 81, 82, 83 - Internal error Restart or replace.
Active
Active
84
F84 - Sensor level
Pick-up amplitude saturated.
Active
Active
97
F97 - Add-on module to old
Replace.
Active
Active
Error code level: W = Warning F = Fatal P = Permanent
ERROR