Doc. No. 06EN003318 Rev. B 3 FlexCOR
TM
Model CMF Series Coriolis Flow Meter
FLUID COMPONENTS, INTL
Direction of Horizontal Flow
Static back pressure / min. 0.1 bar (1.5 psi)
Valve
Valve
Reverse
Forward
Horizontal sensor mounting is recommended for air/gas processes.
A shut-off valve should be installed to facilitate the zero-point adjustment of
the sensor. To avoid elimination of the air form the process, a back pressure
of a minimum of 15. to 3 psi is recommended.
The arrow on the sensor indicates the direction of positive flow (the meter is
able to measure flow in both directions).
The liquid should flow in the direction of the arrow (on the sensor) to avoid
partial emptying of the sensor, especially with low flow.
There should be a check or solenoid valve that closes when the flow is zero
so that the liquid does not flow back and causes partial emptying of the sensor.
Vertical Mounting in Pipe
Vertical mounting
· Gas applications
Vertical mounting
· Liquid applications
The CMF-A is most efficient for gas applications when it is mounted as shown. The
CMF-A is most efficient for liquid applications when it is mounted as shown.
With vertical mounting the orientation of the terminal box is not important, rotation,
however, is not allowed to exceed the stated angles of the sensor.
Vertical Mounting Plug Orientation
Plug
Forward
Reverse
The CMF-A can measure flow in both directions.
Mounting of the Pressure Switch
I
mportant
Read the following information before removing the nipple from the sensor
enclosure.
Avoid getting humidity, liquid or particles into the sensor. It may damage the
meter. Following this procedure is recommended.
1. Leave the sensor in a dry and clean place to acclimate until it obtains
ambient temperature, best at approximately 70°F (20
°C).
2. Be careful when removing the nipple and mounting the pressure switch.
3. Check the pressure switch has been correctly mounted and tightened so
the sealing ring fits tightly. Always replace old sealing rings with new ones
after each disassembly.
Nipple
Ground
(Ex-applications)
Plug
Connecting the Pipe
Important
When connecting/disconnecting the pipe, the cable has to be mechanically
connected in order to prevent liquid from penetrating into the sensor. The
sensor is only IP 65 (dust and splash proof) when the plug is mounted.
Plug