beautypg.com

Cmf-a sensor installation – Fluid Components International CMF Series Guide User Manual

Page 2

background image

FLUID COMPONENTS, INTL

FlexCOR

TM

Model CMF Series Coriolis Flow Meter 2 Doc. No. 06EN003318 Rev. B

CMF-A Mounting Bracket and Position

The mounting brackets supplied with the CMF-A instrument must always be used. The unit must be mounted on a wall or

steel frame (vibration-free).

SENSOR IDENTIFICATION

Remote Mount
Configuration
CMF-B Through F

CMF-A

Nipple

Ground
(Ex-applications)

Bracket
Support

Process connections

Sensor

Plug

CMF-A Plug Orientation

To obtain the optimum performance, the plug should be mounted as shown. The plug can be turned within the angles stated.

Plug

Plug

Plug

Transmitter,
Hazardous Location,
Integral Mount

Transmitter,
Weather Proof,
Integral Mount

Horizontal CMF-A Mounting in Pipe for Liquid or Gas

Horizontal mounting is recommended in low flow liquid so that air bubbles
are easier to remove.

For liquid applications, locate the sensor low in the pipe system in order to

avoid under-pressure in the sensor and consequent air separation in the
liquid. Due to the capillary tube effect, the sensor is not self emptying.

Horizontal mounting

· Gas applications

Horizontal mounting

· Liquid applications

The mounting bracket supplied with the CMF-A
instrument must always be used. The unit must be
mounted on a wall or steel frame (vibration-free).
The plug is used as an indicator for the mounting
position of the sensor. The plug should always be
placed within 5° top center of vertical.

Nipple

Ground
(Ex-applications)

Bracket
Support

Media Input (Preferred)
Process connections

Sensor

Plug

Media Output (Preferred)
Process Connection

CMF-A SENSOR INSTALLATION