Burkert Type 8035 User Manual
Page 57
55
Adjustmentandcommissioning
Type 8025 /8035 Batch
8.7
entering the k factor of the fitting used
The▶device▶determines▶the▶flow▶rate▶in▶the▶pipe▶using▶the▶fitting▶K▶factor.
The▶K▶factor▶of▶the▶fitting▶used▶can▶be▶entered▶here.▶The▶device▶may▶also▶determine▶the▶K▶factor▶using▶a▶teach-in▶procedure:▶see▶the▶detailed▶
operating▶instructions.
The▶device▶will▶use▶the▶new▶K▶factor▶as▶soon▶as▶"SAVE▶YES"▶is▶confirmed▶when▶leaving▶the▶Parameters▶menu.
The▶K▶factor▶of▶the▶fitting▶used▶can▶be▶found▶within▶the▶operating▶instructions▶of▶the▶fitting.
The▶operating▶instructions▶of▶the▶Bürkert▶fittings▶can▶be▶found▶on▶the▶CD▶delivered▶with▶the▶device▶or▶on▶the▶internet▶at▶
www.burkert.com
.
The▶display▶shows▶the▶K▶factor▶of▶the▶fitting,▶
last▶entered▶or▶determined▶using▶a▶Teach-In▶
procedure.
▶
→
Confirm.
K-FACtOR
K=10.000
REtURN
▶
→
Enter▶the▶K▶factor▶(value▶between▶0,01▶and▶99999,9)▶of▶
the▶fitting▶used.
▶
→
Confirm▶the▶displayed▶value.
K=2.85
OUtPUt
VALiD N/Y
Is▶only▶displayed▶if▶the▶K▶factor▶has▶been▶changed.
▶
→
Confirm▶the▶entered▶K▶factor▶or▶not.
Fig. 38: Entering the K factor of the fitting used
▶
→
If▶you▶do▶not▶want▶to▶adjust▶another▶parameter,▶go▶to▶the▶"END"▶parameter▶of▶the▶Parameters▶menu▶and▶press▶
ENTER
▶to▶save▶the▶settings▶or▶
not▶and▶go▶back▶to▶the▶Process▶level.
English