Burkert Type 8035 User Manual
Page 15
13
Technicaldata
Type 8025 /8035 Batch
Power▶supply,▶if▶the▶dosing▶
controller▶is▶energized▶with▶a▶
12-36▶V▶DC▶voltage
supplied▶by▶the▶dosing▶con-
troller▶depending▶on▶the▶position▶
of▶selector▶"SENSOR▶SUPPLY"▶
of▶the▶8025,▶either:
•▶ 5▶V▶DC,▶30▶mA▶max.
•▶ (L+)-12V:▶supply▶voltage▶(L+)▶
of▶the▶dosing▶controller▶minus▶
12▶V▶DC▶(minus▶12,5▶V▶DC▶
max.),▶80▶mA▶max.
•▶ L+:▶supply▶voltage▶(L+)▶of▶
the▶dosing▶controller▶(minus▶
1,5▶V▶DC▶max.),▶140▶mA▶max.
Power▶supply,▶if▶the▶dosing▶
controller▶is▶energized▶with▶a▶
115/230▶V▶AC▶voltage
supplied▶by▶the▶dosing▶con-
troller▶depending▶on▶the▶position▶
of▶selector▶"SENSOR▶SUPPLY"▶
of▶the▶8025,▶either:
•▶ 5▶V▶DC,▶30▶mA▶max.
•▶ (L+)-12V:▶27▶V▶DC▶minus▶
12▶V▶DC▶(minus▶12,5▶V▶DC▶
max.),▶80▶mA▶max.
•▶ L+:▶27▶V▶DC,▶80▶mA▶max.
6.7
electrical connection
Type▶of▶connection
on▶the▶terminal▶blocks▶of▶the▶
electronics▶(and▶through▶cable▶
glands▶for▶the▶compact▶and▶
wall-mounted▶versions)
Cable▶specifications
•▶ cable▶type
•▶ Cross▶section▶of▶wires
•▶ Diameter▶of▶each▶cable▶(for▶
the▶M20x1,5▶cable▶glands▶of▶
the▶compact▶version)▶
▶
▶
▶
•▶ Diameter▶of▶each▶cable▶(for▶
the▶cable▶glands▶M16x1,5▶of▶
the▶wall-mounted▶version)
•▶ shielded
•▶ 0.2▶to▶1.5▶mm
2
•▶ -▶if▶only▶one▶cable▶is▶used▶per▶
M20x1,5▶cable▶gland:▶6▶to▶
12▶mm▶
-▶if▶two▶cables▶are▶used▶per▶
M20x1,5▶cable▶gland:▶4▶mm,▶
with▶the▶supplied▶multi-way▶
seal
•▶ 4▶to▶8▶mm
English