3-6

Brooks

®

Models 5965, 5851EM

Section 3 Operation

Installation and Operation Manual

X-TMF-5965-5851EM-MFC-eng

Part Number: 541B122AAG

September, 2009

Figure 3-5 D-Connector PC Board Jumper Location and Function (Current I/O Version)

B

A

J4

B

J3

A

B

A

J4

N

SS

J2

J7

B

4 - 20 mA

Output

0 - 20 mA

Output

J3

Note: Does not affect voltage output.

Disabled

(Standard)

J2 - RED
SOFT START

Enabled

A

A

A

J7 - GREEN
SETPOINT INPUT SIGNAL TYPE
(Mass Flow Controller only)

CS REV H AND HIGHER.

COMPONENT SIDE SHOWN.

SENSOR

CONNECTOR

R3

R2

R1

R4

VALVE
CONNECTOR

CIRCUIT COMMON TP4

LINEARITY VOLTAGE TP2

VALVE VOLTAGE TP3

TEST POINTS:

SENSOR VOLTAGE TP1

(Multiplied by +100)

A

B

A

J4

B

J3

A

ANTICIPATE

LINEARITY

SPAN

ZERO

ADJUSTMENT POTENTIOMETERS:

0 - 5 VOLTS

4 - 20 mA

J3 & J4 - BLUE
CURRENT OUTPUT RANGE

NOTE:

The 5V Reference is always enabled

and is available on Pin 11.

N

SS

N

SS