beautypg.com

Brooks – Brooks Instrument 5851EM User Manual

Page 11

background image

1-5

Brooks

®

Models 5965, 5851EM

Section 1 Introduction

Installation and Operation Manual

X-TMF-5965-5851EM-MFC-eng
Part Number: 541B122AAG
September, 2009

ELECTRICAL CHARACTERISTICS (continued):

Differential Pressure
Voltage I/O option
Valve orifice sized for any pressure drop between 5 to 50 psi (Minimum
pressure drop depends on gas and range). Refer to Orifice Sizing, Section
4-5.
Current I/O option
Valve orifice sized for any pressure drop between 5 to 1,500 psi (Maximum
pressure drop depends on orifice size). Refer to Orifice Sizing, Section 4-5.

Power Supply Sensitivity
Voltage I/O Option: ±0.09 full scale of percent power supply voltage
variation of full scale.
Current I/O Option: negligible effect over +22.5 to +28 Vdc.

PHYSICAL CHARACTERISTICS:

Materials of Construction

316L, 316L VAR (Vacuum Arc Remelt), and high alloy ferritic stainless
steel.

External seals: Nickel.
Internal seals: Nickel.

Valve Seat: 316L, Viton

®

fluoroelastomers, Buna-N, Kalrez

®

or Teflon

®

.

When 316L or Teflon seat is used, maximum inlet pressure is 50 psig and
minimum controllable flow rate is 3% of full scale.

Dimensions
Refer to Figures 1-2 and 1-3

Process Connections
Integrally welded to body - standard:
1/4" male VCR

TM

3/8" male VCR (3/8" or 1/2" tube)
1/2" ACLIT

TM

1/2" ATW tube stub
CS Downport (5965 Series Only)

Surface Finish
5851EM: 32Ra, Passivated,
Electropolished optional
5965: 7Ra, Electropolished