Dynasonics Fusion Water Flow Meter User Manual
Page 5

5
3. ELECTRICAL CONNECTIONS Route 12-30 volt DC power
and earth ground through one of the FUSION conduit
connections and secure at the 12-30 VDC In, Power Gnd and 
Earth Gnd terminals. Ensure that electrical codes applicable to 
the installation area are followed. An installation drawing 
covering Hazardous Area installations is located in the back of 
this manual. (The FUSION meter contains several output 
options that will be covered in detail in subsequent sections of 
this manual.)
4. APPLY POWER After power is applied to the FUSION flow
meter, the display will illuminate and show the software
version that is running inside of the flow meter as Ux.xx. The 
flow meter will then conduct self-diagnostic, measurement 
and buffering operations during an Auto Adjust countdown 
that is shown on the display. These operations are also 
conducted each time a user exits from the keypad 
configuration menu.
5. OPERATION If the flow meter is started empty, the display
will indicate “EMPTY PIPE” until the flow meter fills with liquid.
If the flow meter is filled with liquid, the flow meter will 
display either DOPPLER or TRANSIT (indicating the mode of 
operation that is supplying the flow readings to the display 
and outputs). 
6. CONFIGURATION There are seven parameters that are
configured in the field on a typical installation.
A. RATE UNIT 
B. TOTAL UNIT 
C. FLOW DIRECTION 
D. OUTPUT <4-20 mA, PULSE> 
E. FULL FLOW 
F. FLOW CUTOFF 
G. RESPONSE 
7. TOTALIZER RESET The flow totalizer can be reset to zero
by simultaneously pressing the MENU and ENTER keys or by 
momentarily connecting the Reset Total In terminal to Power 
Gnd terminal.