Troubleshooting – Dynasonics Fusion Water Flow Meter User Manual
Page 37

37
Troubleshooting
Display does not 
illuminate 
1. Check DC voltage at terminal strip. 
2. Check display ribbon cable. 
3. FUSION meters are equipped with
automatic reset fuses.
Output does not 
match displayed 
flow value
1. Verify that 4-20mA or Rate PULSE is
properly selected in the OUTPUT
selection.
2. Check full flo span value entry—ensure
it matches receiving device.
3. PULSE outputs—verify proper pull-up
jumpers are in place.
4. Utilize 4-20mA Test and PULSE test to
simulate outputs to receiving devices.
Unstable flow 
readings 
1. Increase straight plumbing at meter
inlet.
2. Change response to medium or slow. 
3. Increase damping value. 
Fusion indicates 
flow when none is 
present
1. Verify earth ground connection. 
2. Verify that check valves are operating 
properly—shut manual valve.
3. Perform zero capture—see software
functions.
EMPTY PIPE 
indication 
1. Pipe is not filled with liquid. 
2. Flow is off, but the liquid within the 
sensor is degassing—the error will 
clear when gas clears. 
 
 
 
 
 