TOTAL OP -- Total Operation
POS - Accumulates when flow direction is positive only.
NET - Adds or Subtracts from Total depending upon flow
direction. Applies to transit time mode only.
SET TOTAL -- Preset Totalizer
Enter the value in user selected total units that the flow meter is
to begin totalizing.
PASSWORD -- Change the Security Password
0-9999
By changing the Security Password from 0 to some other value
(any value between 1-9999), configuration parameters will not
be accessible without first entering that value when prompted. If
the value is left at 0, no security is invoked and unauthorized
changes could be made. Access to resetting of the Totalizer is
also protected by this password.
If the FUSION password is lost or forgotten, contact the
Dynasonics factory for a universal password to unlock the meter.
4
‐20CAL -- Calibration of the 4-20 mA Output
IMPORTANT: The 4-20 mA output is factory calibrated and
typically does not require field adjustment.
The 4‐20CAL? entry allows fine adjustments to be made to the
“zero” and span of the 4-20 mA output. Select YES to access
adjustment. To adjust the 4 mA output, a milliammeter or
reliable reference must be connected to the 4-20 mA output.
Procedure:
1. Disconnect the current loop output and connect the
milliammeter in series between the 4-20 mA output and the
recording device.
2. Using the arrow keys, increase the numerical value to
increase the current in the loop to 4 mA. Decrease the value
to decrease the current in the loop to 4 mA. Typical values
range between 40-80 counts.
3. Re-connect the 4-20 mA output circuitry as required.