Dynasonics Fusion Water Flow Meter User Manual

Page 24

background image

24

RATE TST -- Rate Pulse Output Test

Allows a fixed frequency to be output from the rate pulse output.

By incrementing this value, the rate pulse output will transmit
the indicated frequency.
The RATE TEST prompt only appears if the output is
configured for Pulse output. The output will be spanned

between 0 flow and the value entered for Full Flow Rate.

Select the rate tst feature. Use the

and

keys to increment/

decrement output frequency. Verify that at mid-scale, 500 Hz,

that the receiving unit is showing 50% of full flow.

RESPONSE -- Flow Meter Response Time

SLO - Maximum filtering applied

MED - Medium filter applied

FAST- Minimum filtering applied

Utilized to speed up (low dampening) or slow down (high

dampening) the display and output reactions to changes in flow.
1. To select the response of the meter, press the MENU key.
2. Press the

key to display response on bottom display. Press

ENTER key.

3. Press the

key to select FAST, MED or SLO.

4. Press the ENTER key.

5. Press the MENU key to exit.

Advanced Menu Items
Press and hold the MENU key until the meter switches to the
setup mode to access the Advanced menu items. The Advanced

Menu contains all of the items from the Basic menu plus the
following additional parameters:

OP MODE

Auto

TT
Doppler

Selects the operating mode. Auto measures using both transit
time and Doppler. TT locks the unit in transit time mode only.

Doppler locks the unit in Doppler mode only.