Dynasonics Fusion Water Flow Meter User Manual
Page 24

24
RATE TST -- Rate Pulse Output Test
Allows a fixed frequency to be output from the rate pulse output.
By incrementing this value, the rate pulse output will transmit
the indicated frequency.
The RATE TEST prompt only appears if the output is
configured for Pulse output. The output will be spanned
between 0 flow and the value entered for Full Flow Rate.
Select the rate tst feature. Use the
▲
and
►
keys to increment/
decrement output frequency. Verify that at mid-scale, 500 Hz,
that the receiving unit is showing 50% of full flow.
RESPONSE -- Flow Meter Response Time
SLO - Maximum filtering applied
MED - Medium filter applied
FAST- Minimum filtering applied
Utilized to speed up (low dampening) or slow down (high
dampening) the display and output reactions to changes in flow.
1. To select the response of the meter, press the MENU key.
2. Press the
►
key to display response on bottom display. Press
ENTER key.
3. Press the
►
key to select FAST, MED or SLO.
4. Press the ENTER key.
5. Press the MENU key to exit.
Advanced Menu Items
Press and hold the MENU key until the meter switches to the
setup mode to access the Advanced menu items. The Advanced
Menu contains all of the items from the Basic menu plus the
following additional parameters:
OP MODE
Auto
TT
Doppler
Selects the operating mode. Auto measures using both transit
time and Doppler. TT locks the unit in transit time mode only.
Doppler locks the unit in Doppler mode only.