Dynasonics Fusion Water Flow Meter User Manual
Page 21

21
MENU key, press
key to step through setup. Press ENTER to
edit a value. Press ENTER to save a value.
RATE UNIT GPM M3/H BPD LPM
TOTAL UNIT GAL M3 BBL LIT
FLOW DIR
POS NEG
FULL FLOW 0-9999 Rate Unit
FLOW C-OFF 0-999.9 Rate Unit
OUTPUT
NONE 4-20MA PULSE
RESPONSE
FAST MED SLO
Press MENU to exit.
RATE UNT -- Engineering Units for Flow Rate
GPM—Gallons per minute
LPM—Liters
per
minute
BPD—Barrels per days (42 U.S. gallons)
M3/H—Cubic meters per hour
1. To change flow rate units, press the MENU key.
2. Press the
►
key to display RATE UNT on bottom display.
Press ENTER key.
3. Press the
►
key to select new rate unit (four selections
located in the upper right-hand corner of the display.
4. Press the ENTER key.
5. Press the MENU key to exit.
TOTL UNT -- Engineering Units for Flow Totalizer
GAL - U.S. Gallons
LIT - Metric Liters
BBL - Oil Barrels (42 U.S. Gallons)
M3 - Cubic Meters
1. To change flow totalizer units, press the MENU key.
2. Press the ► key to display TOTL UNT on bottom display. Press
ENTER key.
3. Press the ► key to select new totalizer unit.
4. Press the ENTER key.
5. Press the MENU key to exit.